eVol® xr digital analytical syringe is the coupling of two precision devices: a digitally controlled electronic drive and an XCHANGE® enabled analytical syringe.
eVol xr syringes are easily and quickly changed allowing them to be dedicated to individual liquids or methods to prevent possible cross-contamination of reagents.
Syringe volume | 1 mL | 500 µL | 100 µL | 50 µL | 5 µL |
---|---|---|---|---|---|
At 100% (uncalibrated)* | ±0.7% | ±0.7% | ±0.7% | ±1.0% | ±1.0% |
At 100% (calibrated)** | ±0.2% | ±0.2% | ±0.2% | ±0.2% | ±0.2% |
At 10% (calibrated)*** | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ± 1.0% |
* Calibration factor of 1.0000 applied to handle.
** Syringe and handle calibrated at full volume.
*** Syringe and handle calibrated at 10% of capacity. Accuracy based on the 10% dispensed volume closest to zero position.
Note: For best results when performing repeat dispense (one aspiration followed by multiple dispense steps) use a calibration factor of 1.0000.
Syringe volume | 1 mL | 500 µL | 100 µL | 50 µL | 5 µL |
---|---|---|---|---|---|
At 100% | ±0.3% | ±0.3% | ±0.4% | ±0.4% | ±0.5% |
At 10% | ±0.6% | ±0.6% | ±0.7% | ±0.8% | ±1.0% |
Specifications are subject to change without notification.
Please refer to the part number list for more items.