CRS ZPure O₂/H₂O gas filters

CRS gas purifiers and filters

Pure gas is a critical requirement in gas chromatography, spectroscopy, optics, lithography, and numerous other applications in manufacturing and analytical laboratories. The ZPure™ line of filters remove a wide range of contaminants to trace levels.

Features and benefits

  • High efficiency in-line traps with outstanding capacity.
  • High quality activated adsorbents for long purifier life and efficient contaminant removal.
  • Various size and fitting configurations to fit existing installations.
  • Filter lifetime is dependent on quality of incoming gas, and the flow rate.
  • Individually leak-tested.

Recommended applications

It can be used to purify inert gases, He, Ar, N2, and H2, making it ideal for use with GC and GC/MS carrier gas lines. It is also recommended for any application requiring ultra-pure gas free from oxygen and moisture.

Product specifications

ZPure O2/H2O
Volume Function Capacity (nominal-max) Outlet concentration at nominal flow rate Flow rate (nominal-max) Max pressure Dimensions Fittings
130 cc Removes oxygen and water from inert gases 275 - 463 cc oxygen
4.3 - 6.6 g water
Oxygen < 5 ppb
Moisture < 20 ppb
430 cc/min - 1.4 SLPM 68.9 bar / 1000 psi 3.2 cm x 28 cm 1/8" and 1/4" brass and stainless steel compression
475 cc 1020 - 1691 cc oxygen
15.9 - 24.1 g water
1.3 - 4.9 SLPM 3.8 cm x 57 cm
500 cc 1060 - 1780 cc oxygen
16.7 - 25.3 g water
1.6 - 5.2 SLPM 13.8 bar / 200 psi 5 cm x 35 cm
750 cc 1500 - 2671 cc oxygen
25.0 - 38.0 g water
2.5 - 7.7 SLPM 5 cm x 50 cm

1) Oxygen capacity is a function of flow rate - the nominal oxygen capacity is determined using the nominal flow rate. The maximum oxygen capacity is determined using 30% of the nominal flow rate.
2) The nominal water capacity is determined for an inlet impurity level of 200 ppm H2O. The maximum water capacity is determined for an inlet impurity level of 10000 ppm H2O.
3) The nominal hydrocarbon capacity is determined for an inlet impurity level of 500 ppm pentane. The maximum hydrocarbon capacity is determined for an inlet impurity level of 2300 ppm pentane.
4) Nominal flow rate is the recommended flow rate for an estimated gas purifier life of 1 year. This assumes the following inlet impurities: 1 ppm O2, 1 ppm H2O, and 1 ppm HC (C5 and heavier). The maximum recommended flow rate is recommended for intermittent use only.

More

Please refer to the part number list for more items.

For technical information such as manuals and MSDS please visit chromres.com/resources.